Small Sensor Array Offers Easier Way to Image Brain
Microextrusion Tubing Advances Help Push Medical Device Boundaries
Adapting Prototyping to Meet Rapid Advances in Medical Device Designs
Designing for Mechanical and Signal Integrity in Handheld Medical Treatment Applications
New Device Helps Diagnose and Treat Neurodegenerative Diseases
ADHESIVES: The Importance of CTE for Assembly Reliability
Hydrogel Material Improves Success of Transplanting Islet Cells
Skin-worn Wearable Devices: Tomorrow’s Vision, Today’s Realities
New 3D Printing Method Promises New Era for Medical Implants
Features

Collaborating with a Medical Electronics OEM

Fig. 3 – A second stage of ultrasonic cleaning is conducted using de-greaser machines.
Multiple cleaning solutions evaluated in close collaboration with a medical electronics OEM. Standard cleaning solutions were the most logical choices, especially with an in-house de-greaser machine to perform ultra cleaning. However, those didn’t produce the expected results, and the boards were not extremely clean. At this point, a DOE was initiated.

A six step-by-step method was used in this DOE to arrive at a satisfactory solution. It begins with the method normally used, but with extra scrutiny to determine if anything was missed.

Step one: The first step was to subject the flex circuit to a de-ionized (DI) water cleaning. In this case it didn’t resolve the issue. Batch cleaning then followed.

Step two: An Ionograph test was performed. However, cleaning results weren’t as effective and as pure as expected.

Step three: The assembly team returned to the drawing board. Brain storming followed and other alternatives were considered.

Step four: Discussions about possible solutions were conducted between the OEM customer and assembly personnel. Another possible answer was agreed to keeping in mind the failed approaches thus far in this step-by-step process.

Step five: The methodology and specifications were changed based on out-of-the-box thinking, and the DOE was re-done.

Step six: It’s well understood that the DOE solution is sometimes within uncharted territory. But both the OEM and assembly personnel continued until the right DOE solution was reached.

DOE solutions arise from special EMS provider in-house methodologies and techniques. The proper solution may not be immediately obvious. But when certain, well-proven experienced thinking and methods are implemented, the correct answer can be reached.

Nitrogen-purged baking at a 2,500 parts per million for two hours at 120°C became the ultimate cleaning solution to rid the medical electronics flex circuit of the bubbles created during the assembly process. The medical electronics flex circuit was exposed to the ambient environment, which has oxygen content and moisture. Nitrogen-purged baking got rid of the oxygen content from the atmosphere to eliminate the bubbles.

Other Types of DOEs

Fig. 4 – Chip-on-board or CoB.
Types of DOEs can include:

  • Crystallizing futuristic technologies;
  • Resolving anticipated technology issues in subsequent product generations;
  • Determining long-term product reliability; and
  • Evaluating different PCB design and assembly practices to improve their product yield.

Following are several examples of issues that can befall medical electronics PCB products. For example, an OEM customer wants to know how their product would age in 5, 10, and 20 years. However, they want to collect that data in a matter of days and weeks to learn the level of failure. If an integrated circuit (IC) failure occurs, they want to know where the weakest link of their product is. That’s half the deal; the other half is determining the manufacturing steps that need to be taken to address that weak link in the overall failure and address it in the next revision so the product doesn’t fail.

Meanwhile, a chipmaker has come up with using a new alloy to make the lead termination. They want to test it and learn how viable it is in manufacturing. The selected EMS provider would use the IC with the new lead terminations and run a multiple set of experiments. Different conditions are used to ascertain how their new alloy works under the different set of conditions. Those include temperature cycles, heat cycle, reflow conditions, soldering parameters, printing adhesion to the surface, and several other conditions.

A third example is a chip-on-board or CoB, as shown in Figure 4. The OEM wants to use a different kind of flux and/or alloy for the solder paste. Then the OEM customer wants to learn how effectively that CoB will undergo assembly processes. A similar DOE is evaluating an FR4 version of a metal core board. In this case, the OEM is interested in how and how much heat transfers when this board is subjected to different temperature cycles, thermal profiles, and soak levels.

Each DOE is unique in the sense that it is characterized by an OEM’s special requirements. However, all DOEs share a variety of different PCB design and assembly aspects and stages such as board finishes, temperature cycles, thermal profiles, solder pastes, different fluxes, chemical agents, and so on.

Considerations

A medical electronics OEM seeking the right EMS provider to perform its DOEs should choose an EMS company that has latitude, flexibility, and most important, technical experience. It’s helpful if the EMS provider has a strong and entrepreneurial mindset to take on new projects and new challenges. Persistence is the linchpin for accomplishing successful DOEs. The understanding between the EMS provider and OEM in these instances is the fact the project at hand is floating in unchartered waters. All the answers aren’t there. But the critical ingredients are available, and those are tenacity, knowledge, basic information, and some idea of the direction that needs to be taken.

This article was written by Zulki Khan, President and Founder, NexLogic Technologies, Inc., San Jose, CA. For more information, Click Here

« Start Prev 1 2 Next End»